Standard ID | JJG 926-2015 (JJG926-2015) | Description (Translated English) | Verification Regulation of Record Pressure Gauges, Pressure Vacuum Gauges and Vacuum Gauges | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A53 | Classification of International Standard | 17.100 | Word Count Estimation | 15,123 | Date of Issue | 2015-01-30 | Date of Implementation | 2015-07-30 | Older Standard (superseded by this standard) | JJG 926-1997 | Quoted Standard | JJF 1008-2008; JJG 52-2013; GB/T 1226-2010 | Regulation (derived from) | The State Administration of Quality Supervision, Inspection and Quarantine Announcement 2015 No. 19 | Issuing agency(ies) | State Administration of Quality Supervision, Inspection and Quarantine | Summary | This Standard applies to recording pressure gauges, pressure vacuum gauge initial verification and vacuum table, and use the subsequent verification check. |
JJG 926-2015
Verification Regulation of Record Pressure Gauges, Pressure Vacuum Gauges and Vacuum Gauges
State verification procedures People's Republic of China
Recording pressure gauges, pressure vacuum gauge and vacuum gauge
RecordPressureGauges, PressureVacuumGaugesandVacuumGauges
Issued on. 2015-01-30
2015-07-30 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Recording pressure gauges, pressure
Vacuum gauge and vacuum meter testing procedures
Replacing JJG 926-1997
Focal point. the National Technical Committee of pressure measurement
Main drafting unit. Shanghai Institute of Measurement and Testing
China Testing Technology Research Institute
Zhejiang Institute of Metrology
Participated in the drafting unit. Xinjiang Institute of Measurement and Testing
Shanghai Pudong New Area of the Metrology and Quality Inspection
The measurement procedure commissioned the National Technical Committee is responsible for interpreting pressure
The main drafters of this procedure.
Tu Li Meng (Shanghai Institute of Measurement and Testing)
Gan Rong (China Testing Technology Research Institute)
Cai Fei (Zhejiang Institute of Metrology)
Drafters participate.
Zhuo China (Xinjiang Institute of Measurement and Testing)
Li Yan Ling (China Testing Technology Research Institute)
Ling Ming (Shanghai Pudong New Area Metrology and Quality Inspection Institute)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2. References (1)
3 terminology and units of measurement (1)
4 Overview (1)
5 metering performance requirements (1)
5.1 Accuracy class and maximum permissible error (1)
5.2 Indication error (2)
5.3 Return error (2)
5.4 Displacement tap (2)
5.5 stylus movement smooth (2)
5.6 Error when walking (2)
6 General Requirements (2)
6.1 Appearance (2)
6.2 stylus and record line (3)
7 control of measuring instruments (3)
7.1 test conditions (3)
7.2 Verification Project (4)
7.3 Verification Methods (4)
7.4 test results processing (6)
7.5 test cycle (6)
Appendix A recording pressure gauge test recording format (7)
Appendix B recording pressure gauge calibration certificate/test results in the notice page format (8)
B.1 recording pressure gauge calibration certificate test results page format (8)
B.2 recording pressure gauge test results test results notice page format (9)
introduction
JJF1002 "national verification procedures write rules", JJF1001 "common measurement terms and definitions"
JJF1059 "Evaluation and Expression of Uncertainty in Measurement" series together constitute the basic norms of the regulations revision.
Revised procedures combined with China's national conditions, the use of the International Organization of Legal Metrology (OIML) International Recommendation R112
"Indicates recording type and pressure gauge with an elastic element, vacuum gauge, pressure - vacuum gauge (General Instrument)" part
Content is reserved JJG 926-1997 "recording pressure gauges, pressure vacuum gauge and vacuum gauge test procedures" line of
Revision under the premise of effective content. This procedure compared with JJG 926-1997, in addition to editorial changes, the main
Technical changes are as follows.
--- Press JJF1002-2010 "national verification procedures write rules" requirement, increase the introduction
Content;
--- Regulations covering the content JJG 926-1997 apply to all records pressure gauges, pressure vacuum
Verification table and vacuum table;
--- Delete the contents of the measuring range upper limit of the range;
--- Improve the content of the overview;
--- Refine the test conditions;
--- Straighten out the order of test items and test methods;
--- Remove the "0.25, 0.4, 0.6" Recording pressure gauge accuracy classes;
--- An increase of "90% to 100%" Pressure measuring range tolerance requirements;
--- Added "tap shift", "smooth stylus movement";
--- Added "test certificate/test results within the notice page and Format" appendix.
The regulations previously issued as follows.
--- JJG 926-1997.
Recording pressure gauges, pressure vacuum gauge and vacuum meter testing procedures
1 Scope
The order applies to recording pressure gauges, pressure vacuum gauge and vacuum gauge (hereinafter referred to as gauge) of the first seizure
Set, and use the subsequent verification check.
2. References
This procedure references the following documents.
JJF1008-2008 "Pressure measurement terminology and definitions"
JJG 52-2013 "elastic element type general pressure gauges, pressure vacuum gauge and vacuum gauge"
GB/T 1226-2010 "General gauges"
For dated references, only the dated edition applies to this procedure; undated reference documents
Member, the latest edition (including any amendments) applies to this procedure.
3 terminology and units of measurement
3.1 Terms
3.1.1 record pressure gauge recordpressuregauge
Amount of pressure can be continuously recorded on the recording paper gauge. [JJF1008-2008 defined 4.24]
3.1.2 recording paper recordpaper
Continuous recording sheet pressure magnitude and specific functions of travel time.
3.1.3 Error traveltimeerror go
Finish the trip within the stipulated time and the time difference with the standard time.
3.2 units of measurement
Legal units of measurement used to gauge Pa (Pascal), or its decimal multiples units. kPa, MPa
Etc; time unit of measure as the legal units of measurement s (seconds) or min (minute) and h (hours).
4 Overview
Pressure gauge is mainly used for liquid, gas and vapor measurement.
Gauge works by using elastic sensing element (such as a diaphragm spring tube or the like) to produce bomb under pressure
Plastic deformation - displacement, size, shape and function of its variable pressure into a linear relationship, by means of connecting rod drive
Mechanism (movement) stylus driven by a stylus on a moving recording paper record of the measured pressure.
5 metering performance requirements
5.1 accuracy class and the maximum permissible error
Pressure gauge accuracy classes and maximum permissible errors shall comply with Table 1.
......
|